Test 3. Evaluation Of ELISA PlateAim Materials and methods Materials: ELISA paltes Reagents Instrument Method: 96 well ELISA plates were tested by classic ELISA. Briefly, coat the plates with 8ug/ml human IgG. Then the plates were blocked and added with goat-anti-human IgG-HRP. TMB was added as substrate. sulphuric acid was used to stop the reaction. The plates were read at 450-630nm. The experiments were repeated 5 times. Results The data were shown in table 5. Table 5 absorbance of ELISA plates at 450-630nm:
* Significant difference (P<0.05) The CV value at 450-630nm was 3.4%. Conclusion From the above data we can conclude that the binding capacity of biousing 96-well ELISA plate is better than that of Brand C. The CV value is below average value 5%. Maybe you also want to visit:
|